Product Details:
|
Accuracy: | 5% | Acceleration Amplitude(RMS): | 9.81m/s2(1g)±3% |
---|---|---|---|
Output Waveform: | Sine Waveform | Waveform Distortion: | 5% |
Maximum Load: | 120Gram | Working Temperature: | 0-55º C |
High Light: | ndt testing equipment,non destructive testing machine |
HG5003 Vibration Calibratorcan be used to calibrate small vibration transducers and also vibration measuring instruments
HG5003 Vibration Calibrator
General Description
HG5003 Vibration Calibrator can be used to calibrate small vibration transducers and also vibration measuring instruments and vibration monitoring systems quickly. Standard sine signals with fixed frequency 159.2Hz and fixed amplitude is generated from the calibrator.
HG5003 Vibration Calibrator combines sine signal generator, power amplifier, standard transducer and shaker in one and features a smaller volume, high accuracy and easy operation. It can be used either in laboratory or on site.
Operation
1
Fix M5 screw, accelerometer on the shaker table of the calibrator
2
Press the only switch at the calibrator to start the vibration.
3
Compare the reading of your vibration meter before the shaker power off after 50sec.
4
The LED will turn from GREEN to ORANGE when battery low. It means you need to recharge the battery for 15 hours.
Contact Person: Ms. Shifen Yuan
Tel: 8610 82921131,8613910983110
Fax: 86-10-82916893